Televac AN 3009: MM200 to PC Interfacing and Communications

Description

This application note explains the process for interfacing and communicating with an MM200 from a Windows PC. The serial communications program Tera Term must be installed prior to following these instructions. If you don’t have Tera Term installed, click here to install the latest version from GitHub.

Instructions

1. Connect the MM200 serial port on the back of the unit to your PC, either using a serial to USB converter or a serial port on your PC. Note that a straight through D-sub 9 pin cable should be used.

2. Open the Start Menu.

Application Note 3009 | Figure 1

3. Open the Control Panel.

Application Note 3009 | Figure 2

4. Open the Hardware and Sound.

Application Note 3009 | Figure 3

5. Open the Device Manager.

Application Note 3009 | Figure 4

6. Expand the Ports (COM & LPT) drop down.

7. Determine the COM port assignment for your MM200, in this example the MM200 is assigned to COM3.

Application Note 3009 | Figure 5

8. Determine the COM port assignment for your MM200, in this example the MM200 is assigned to COM3.

9. Open Tear Term.

10. Open the Setup menu and select the Serial port… option

Application Note 3009 | Figure 6

11. Change the Port setting to the COM port for your MM200, in this example it is COM3.

12. Ensure that the remainder of the settings match what is shown below, then select OK. Note that this example uses the default baud rate of 9600.

  1. Baud rate: 9600
  2. Data: 8 bit
  3. Parity: none
  4. Stop: 1 bit
  5. Flow control: Xon/Xoff
  6. Transmit delay: 0 msec/char, 0 msec/line
Application Note 3009 | Figure 7

13. Open the Setup menu and select the Terminal… option.

Application Note 3009 | Figure 8

14. Ensure that the settings are the following, then select OK:

  1. Receive: CR+LF
  2. Transmit: CR+LF
  3. Local echo checked
Application Note 3009 | Figure 9

15. You should now be able to communicate with the MM200. As an example, send the following command to read the setup information for the unit: “ST<cr>” (not including the quotations, <cr> is equivalent to pressing the enter key).

16. If your unit is in burst mode, not all commands will be usable. To exit burst mode, use the following command (<cr> is equivalent to pressing the Enter key):

  • BF<cr>

17. If you want to re-enter burst mode, use the following command:

  • BN<cr>

Related Products

Televac®️ MM200 Vacuum Controller

MM200 Vacuum Controller

• Customer configurable design
• RS-232/RS-485 digital communications
• Supports up to 8 relay set points
• Selectable units of measurement