Beschreibung
Historically the 2A thermocouple sensor had a measurement range up to only 20 Torr and the 4A convection sensor was used to measure pressures up to 1000 Torr. Software and drive signal enhancements incorporated into the MX2A active thermocouple gauge now allow the 2A thermocouple sensor to measure pressures up to 1000 Torr. This has raised questions about the differences between the MX2A thermocouple active gauge and the MX4A active convection active gauge as almost all of their specifications are now identical. This document outlines the differences between the two active gauges.
Unterschiede zwischen dem MX2A und dem MX4A
Beschreibung | MX2A | MX4A |
Sensor Typ | Thermoelement | Konvektion |
Filament-Durchmesser | Dicker | Verdünner |
Beständigkeit gegen Verschmutzung | Höher | Unter |
Reaktionszeit | <2 seconds | <1 second |
Genauigkeit | ±30% (10 bis 100 Torr) | ±10% (10 bis 100 Torr) |
Materialien, die einem Vakuum ausgesetzt sind | NiMnSi-Legierung, CuNiMn-Legierung | NiCr-Legierung, CuNi-Legierung |
Kosten | Niedrigste | Niedrig |
Alle anderen Spezifikationen für das MX2A und MX4A sind gleichwertig.
Verwandte Produkte
MX2A Thermoelement-Aktiv-Vakuum-Messgerät
- 1*10-4 Torr bis zu 1000 Torr
- Leicht ablesbares farbiges OLED-Display
- Leicht austauschbare Sensoren
MX4A Konvektions-Aktiv-Vakuummessgerät
- 1*10-4 Torr bis 1000 Torr
- Leicht austauschbare Sensoren
- Verschmutzungsresistent