Televac MP7FR Kaltkathode

Not recommended for new designs.
Replaced by Televac® MX7B and MX7M Active Gauge.

Part Numbers: 2-7970-152, 2-7970-352


Merkmale und Vorteile:

  • Kompakte Bauweise
  • In jeder Orientierung montierbar
  • Minimales magnetisches Streufeld
  • Überlegene Leistung von Gerät zu Gerät
  • Ausgezeichnete Kundenbetreuung

The Televac® MP7FR Active UHV Gauge utilizes an easily replaceable Televac® 7F double inverted magnetron cold cathode vacuum sensor. It has a variety of features including a wide range of measurement from 1*10-11 Torr to 1*10-2 Torr, a programmable set point, a factory settable analog output, and an LED display. This unit indirectly measures absolute pressure by ionizing gas in the presence of a magnetic field. Unlike hot ion sensors, this unit is resistant to inrushes of gas. The design allows for easy disassembly for replacement of the sensor.

Betriebsbereich 1*10-11 Torr bis 1*10-2 Torr
Kommunikation N/A
Analoger Ausgang 0 V bis 10 V DC
Programmierbare Sollwerte 1
Sollwert-Typ Relais
Versorgungsspannung 24 V DC
Maximale Leistung 6 W
Kalibriermedium Trockene Luft oder Stickstoff
Überdruck 150 psi
Auflösung des digitalen Ausgangs 2 signifikante Ziffern mit Exponent
Betriebstemperatur 0 °C bis 50 °C
Lagertemperatur -20 °C bis 60 °C
Ausheiztemperatur 150 °C (Elektronik entfernt)
400 °C (Elektronik und Magnete entfernt)
Anzeige Lesbarer Abstand 3 m (10 ft)
 
Genauigkeit
1*10-7 Torr bis 1*10-2 Torr ±0,2 Dekade
Analoger Ausgang ±5 mV
Wärmebehandlungs- und Vakuumöfen
Kryogenik und Industriegase
Dünnschichtabscheidung und Beschichtungsprozesse
Halbleiter
Gefriertrocknung und Lebensmittelverpackung
Lastverriegelung
Pharmazeutische Produktprozesse