Part Numbers: 2-7980-031, 2-7980-032, 2-7980-052
Características y beneficios:
- Un medidor de presión de vacío de amplio rango muy compacto
- Pantalla digital de fácil lectura del vacuómetro LED
- Un vacuómetro de excelente resolución y repetibilidad
- Rendimiento superior de unidad a unidad entre los vacuómetros
- ISO 17025 accredited and NIST traceable calibrations optional
- Excelente apoyo al cliente para los vacuómetros
The Televac® CC-10 Wide Range Active Vacuum Gauge (also called an analog vacuum gauge, digital vacuum gauge, or vacuum pressure gauge) utilizes two internal sensors to measure full range vacuum from UHV (10-9 Torr) to above atmosphere (103 or 1000 Torr). This vacuum gauge has a variety of features including RS-485 communications, three programmable set points, and a selectable analog vacuum pressure gauge output.
La brillante pantalla del vacuómetro digital LED hace que sea fácil de leer a distancia. Las opciones de este medidor de presión de vacío de amplio rango se pueden configurar desde cuatro controles táctiles en el frente de la unidad o a través de comunicaciones RS-485. Este vacuómetro ofrece una amplia selección de accesorios, un uso sencillo y una fácil sustitución del sensor, lo que lo convierte en una excelente opción para una gran variedad de aplicaciones de vacío.
Medidor de Vacío Activo de Amplio Alcance CC-10
D-Sub 15, 15 Conductor Cable (Flying Leads)
Fuente de alimentación de la pared
Piezas de repuesto
Rango de operación | 1*10-9 Torr a 1000 Torr |
Comunicaciones | RS-485 |
Salida analógica | 0 V a 10 V lineal/logarítmico |
Puntos de ajuste programables | 3 relevos |
Voltaje de suministro | 22 V DC a 26 V DC |
Máxima potencia | 8 W |
Medio de calibración | Aire seco o nitrógeno |
Sobrepresión | 150 PSI |
Resolución de la salida digital | 2 dígitos significativos con exponente |
Resolución de la salida analógica | 16 bits |
Temperatura de funcionamiento | 0 °C a 50 °C |
Temperatura de almacenamiento | -40 °C a 85 °C |
Temperatura de cocción | 110 °C (electronics removed) |
Tiempo de respuesta | ≤500 ms |
Display Readable Display | 4 m (13 pies) |
Precisión | |
10-9 Torr to 10-1 Torr | ±0,2 de década |
10-1 Torr to 200 Torr | ±15% of reading |
200 Torr to 1000 Torr | ±5% de la lectura |
Salida analógica | ±10 mV |
Pin | Descripción |
1 | Analog Out (Ground) |
2 | RS-485 A (-) |
3 | SP3 Out |
4 | SP2 Out |
5 | SP1 Out |
6 | Zero adjust input (signal in) |
7 | EXT Vcc (signal in) |
8 | Supply Voltage (+) |
9 | Analog Out (0 to 10) V |
10 | RS-485 B (+) |
11 | SP2, SP3 COM |
12 | SP1 COM |
13 | HV inhibit signal (signal in) |
14 | Atm adjust input (signal in) |
15 | Supply Voltage (-) |
<STX>0R1<CR> | Read units |
<STX>0R2<CR> | Read SP1L and SP1H |
<STX>0R5<CR> | Read analog output set parameter |
<STX>0W1XXXX<CR> | Set pressure unit |
<STX>0W2XXXXXXXX<CR> | Set SP1L and SP1H |
<STX>0C1<CR> | Crystal gauge atm adjustment |
<STX>0S1<CR> | Read pressure data |
<STX>0S2<CR> | Read error status |
Assume address 0
Full list of commands is available in the instruction manual.
The CC-10 has three set points. All three are optically coupled solid state relays that are rated for 0.12 A at 50 V DC. Each set point has hysteresis characteristics and consists of two set values: upper threshold and lower threshold. LEDs on the front panel display the status of each set point. The high and low threshold values of each of the three set points can be set independently.
Molibdeno |
Glass |
Copper |
Stainless Steel |
Chromium |
Solder |
Fluorocarbon Elastomer |
Cuarzo |
Enclosure Material | Metallized plastic |
Conexiones eléctricas | D-Sub 15-pin male |
El peso con el sensor | 698 g (1.54 lbs) (NW25 Flange) |
El peso sin el sensor | 355 g (0.78 lbs) |
CE certified to EN 61010-1, EN 61236-1 |
RoHS compliant |
IP40 |
ISO 17025 accredited calibration optional |
Calibración trazable NIST opcional |
Documentación del producto
CC-10 Wide Range Datasheet |
CC-10 Wide Range Instruction Manual |
Televac®️ Vacuum Calibration Tolerances |
3D Models (STEP Files)
2-7980-031 – CC-10 NW25/KF25 Stainless Steel |
2-7980-032 – CC-10 NW40/KF40 Stainless Steel |
2-7980-052 – CC-10 CF40/2.75″ CF-F Stainless Steel |
Notas de aplicación
Televac® Application Note 3015: Recommended Practices for Vacuum Calibration |
Herramientas
Convertidor de medición de vacío Televac® |
Web Serial Terminal |
Artículos
Vacuum Measurement Using Modern Cold Cathode Technology |
Brochures
Televac®️ Vacuum Measurement Product Catalog |
CC-10 Wide Range Active Vacuum
Gauge Quantity Discounts
Cantidad | Discount |
1 | List |
5 | 2% |
10 | 5% |
15 | 7% |
20 | 10% |
25 | 13% |
30+ | 15% |
D-Sub 15 Cable Quantity Discounts
Cantidad | Discount |
1 | List |
50 | 5% |
75 | 10% |
100+ | 15% |