Description

Historically the 2A thermocouple sensor had a measurement range up to only 20 Torr and the 4A convection sensor was used to measure pressures up to 1000 Torr. Software and drive signal enhancements incorporated into the MX2A active thermocouple gauge now allow the 2A thermocouple sensor to measure pressures up to 1000 Torr. This has raised questions about the differences between the MX2A thermocouple active gauge and the MX4A active convection active gauge as almost all of their specifications are now identical. This document outlines the differences between the two active gauges.

Différences entre le MX2A et le MX4A

Description MX2A MX4A
Type de capteur Thermocouple Convection
Diamètre du filament Plus épais Diluant
Résistance à la contamination Plus haut En bas
Temps de réponse <2 seconds <1 second
Précision ±30% (10 à 100 Torr) ±10% (10 à 100 Torr)
Matériaux exposés au vide Alliage NiMnSi, alliage CuNiMn Alliage NiCr, alliage CuNi
Coût Le plus bas Faible

Toutes les autres spécifications pour le MX2A et le MX4A sont équivalentes.