Description
Historically the 2A thermocouple sensor had a measurement range up to only 20 Torr and the 4A convection sensor was used to measure pressures up to 1000 Torr. Software and drive signal enhancements incorporated into the MX2A active thermocouple gauge now allow the 2A thermocouple sensor to measure pressures up to 1000 Torr. This has raised questions about the differences between the MX2A thermocouple active gauge and the MX4A active convection active gauge as almost all of their specifications are now identical. This document outlines the differences between the two active gauges.
Différences entre le MX2A et le MX4A
Description | MX2A | MX4A |
Type de capteur | Thermocouple | Convection |
Diamètre du filament | Plus épais | Diluant |
Résistance à la contamination | Plus haut | En bas |
Temps de réponse | <2 seconds | <1 second |
Précision | ±30% (10 à 100 Torr) | ±10% (10 à 100 Torr) |
Matériaux exposés au vide | Alliage NiMnSi, alliage CuNiMn | Alliage NiCr, alliage CuNi |
Coût | Le plus bas | Faible |
Toutes les autres spécifications pour le MX2A et le MX4A sont équivalentes.
Produits connexes
Jauge à vide active pour thermocouple MX2A
- 1*10-4 Torr jusqu'à 1000 Torr
- Écran OLED couleur facile à lire
- Des capteurs faciles à remplacer
Jauge à vide active à convection MX4A
- 1*10-4 Torr à 1000 Torr
- Des capteurs faciles à remplacer
- Résistant à la contamination