Télévac AN 3009 : Interfaces et communications MM200 vers PC

Description

This application note explains the process for interfacing and communicating with an MM200 from a Windows PC. The serial communications program Tera Term must be installed prior to following these instructions. If you don’t have Tera Term installed, click here to install the latest version from GitHub.

Instructions

1. Connect the MM200 serial port on the back of the unit to your PC, either using a serial to USB converter or a serial port on your PC. Note that a straight through D-sub 9 pin cable should be used.

2. Ouvrez le menu Démarrer.

3. Ouvrez le Panneau de configuration.

4. Open the Hardware and Sound.

5. Ouvrez le Gestionnaire de périphériques.

6. Développez le menu déroulant Ports (COM & LPT).

7. Determine the COM port assignment for your MM200, in this example the MM200 is assigned to COM3.

8. Determine the COM port assignment for your MM200, in this example the MM200 is assigned to COM3.

9. Open Tear Term.

10. Open the Setup menu and select the Serial port… option

11. Change the Port setting to the COM port for your MM200, in this example it is COM3.

12. Ensure that the remainder of the settings match what is shown below, then select OK. Note that this example uses the default baud rate of 9600.

  1. Taux de baud : 9600
  2. Données : 8 bits
  3. Parité : aucune
  4. Stop : 1 bit
  5. Contrôle des flux : Xon/Xoff
  6. Retard de transmission : 0 msec/char, 0 msec/ligne

13. Open the Setup menu and select the Terminal… option.

14. Ensure that the settings are the following, then select OK:

  1. Recevoir : CR+LF
  2. Transmettre : CR+LF
  3. Vérification de l'écho local

15. You should now be able to communicate with the MM200. As an example, send the following command to read the setup information for the unit: “ST<cr>” (not including the quotations, <cr> is equivalent to pressing the enter key).

16. If your unit is in burst mode, not all commands will be usable. To exit burst mode, use the following command (<cr> is equivalent to pressing the Enter key):

  • BF<cr>

17. If you want to re-enter burst mode, use the following command:

  • BN<cr>