Accuracy Specification and Vacuum Test Points with Tolerances
Contents
Section 1 – Description
Section 2 – Comparison Table of Calibration Options
Section 3 – 2A, 2A VacuMini, 2A NASA Mini Part 1
Section 4 – 2A, 2A VacuMini, 2A NASA Mini Part 2
Section 5 – 4A Part 1
Section 6 – 4A Part 2
Section 7 – 7B, 7E, 7M, 7F, 7FC, 7FCS Part 1
Section 8 – 7B, 7E, 7M, 7F, 7FC, 7FCS Part 2
Section 9 – CC-10
Section 10 – 1E, 1F
Section 11 – CDG (Capacitance Diaphragm Gauge, Capacitance Manometer)
Section 1 – Description
General: This document gives a description of the calibration options available from Televac® as well as the vacuum calibration test points and tolerances for various vacuum sensors, vacuum controllers, and active vacuum gauges offered by Televac®. All values are in Torr and calibrations are done in nitrogen gas unless otherwise stated. Televac® reserves the right to change vacuum test points and tolerances for any product. Any modification to any unit without approval from Televac® voids the unit’s calibration certificate or certificate of compliance. Additional custom test points are available upon request.
Description of ISO 17025 accredited calibration: The Fredericks Company is ISO 17025 accredited for vacuum and pressure calibration in the range of 1E-6 Torr to 1000 Torr. Units are vacuum calibrated (on a vacuum stand) in nitrogen gas at various points (gauge dependent) against reference standards traceable to NIST, using systems that are accredited to ISO 17025. This includes incoming (as found) data, outgoing (as left) data after calibration adjustments, and the reference standard uncertainty at each point. A statement of conformity using a simple decision rule is included. The conformity decision is made by comparing the measurement results to the calibration limits, defined by form televac_tolerances, without including measurement uncertainty.
Description of NIST traceable calibration: The Fredericks Company offers NIST traceable calibration, where units are vacuum calibrated (on a vacuum stand) in nitrogen gas at various points (gauge dependent) against reference standards traceable to NIST. This includes incoming (as found) data, outgoing (as left) data after calibration adjustments, but does not include the reference uncertainty at each point (ISO 17025 accredited calibration does include uncertainties).
Description of factory (standard) calibration: The Fredericks Company offers standard (factory) calibration, where units are electrically verified using electrical equipment only. The verification includes checking that the units provide vacuum readings within a specified tolerance when a specific electrical signal is applied. This is not a vacuum calibration, and does not guarantee that a sensor will read accurately when connected to a controller or active gauge. This is because sensor contamination and sensor to sensor mechanical variability cannot be accounted for with an electrical verification.
Section 2 – Comparison Table of Calibration Options
Description | Factory (Standard) | NIST Traceable | ISO 17025 Accredited |
Electrical verification | Yes | No | No |
Vacuum comparison calibration to NIST traceable references | No | Yes | Yes |
Reference standard uncertainties included | No | No | Yes |
ISO 17025 accredited calibration | No | No | Yes |
Section 3 – 2A, 2A VacuMini, 2A NASA Mini Part 1
Applicable Controllers and Active Gauges MX200, MM200, MC300, MV2A, Compact 2A, VacuGuard™, B2A Portable, MP2AR
Factory (Standard) Calibration
Test Point | Lower Tolerance | Upper Tolerance |
0.0E+0 | -1.0E-3 | 1.0E-3 |
1.0E-1 | 9.0E-2 | 1.1E-1 |
1.0E+0 | 8.0E-1 | 1.2E+0 |
NIST Traceable Calibration
Test Point | Lower Tolerance | Upper Tolerance |
0.0E+0 | -1.0E-3 | 1.0E-3 |
1.0E-2 | 9.0E-3 | 1.1E-2 |
2.0E-2 | 1.8E-2 | 2.2E-2 |
5.0E-2 | 4.5E-2 | 5.5E-2 |
1.0E-1 | 9.0E-2 | 1.1E-1 |
2.0E-1 | 1.6E-1 | 2.4E-1 |
5.0E-1 | 4.0E-1 | 6.0E-1 |
1.0E+0 | 8.0E-1 | 1.2E+0 |
NIST Traceable GE Aviation Specification P10TF3 Compatible Calibration
Test Point | Lower Tolerance | Upper Tolerance |
0.0E+0 | -1.0E-3 | 1.0E-3 |
1.0E-2 | 9.0E-3 | 1.1E-2 |
2.0E-2 | 1.8E-2 | 2.2E-2 |
5.0E-2 | 4.5E-2 | 5.5E-2 |
1.0E-1 | 9.0E-2 | 1.1E-1 |
2.0E-1 | 1.7E-1 | 2.3E-1 |
5.0E-1 | 4.3E-1 | 5.8E-1 |
1.0E+0 | 8.5E-1 | 1.2E+0 |
ISO 17025 Accredited Calibration
Test Point | Lower Tolerance | Upper Tolerance |
1.0E-3 | 0.0E-3 | 2.0E-3 |
1.0E-2 | 9.0E-3 | 1.1E-2 |
2.0E-2 | 1.8E-2 | 2.2E-2 |
5.0E-2 | 4.5E-2 | 5.5E-2 |
1.0E-1 | 9.0E-2 | 1.1E-1 |
2.0E-1 | 1.6E-1 | 2.4E-1 |
5.0E-1 | 4.0E-1 | 6.0E-1 |
1.0E+0 | 8.0E-1 | 1.2E+0 |
ISO 17025 Accredited GE Aviation Specification P10TF3 Compatible Calibration
Test Point | Lower Tolerance | Upper Tolerance |
1.0E-3 | 0.0E-3 | 1.0E-3 |
1.0E-2 | 9.0E-3 | 1.1E-2 |
2.0E-2 | 1.8E-2 | 2.2E-2 |
5.0E-2 | 4.5E-2 | 5.5E-2 |
1.0E-1 | 9.0E-2 | 1.1E-1 |
2.0E-1 | 1.7E-1 | 2.3E-1 |
5.0E-1 | 4.3E-1 | 5.8E-1 |
1.0E+0 | 8.5E-1 | 1.2E+0 |
Section 4 – 2A, 2A VacuMini, 2A NASA Mini Part 2
Applicable Controllers and Active Gauges
MX2A
Factory (Standard) Calibration
Test Point | Lower Tolerance | Upper Tolerance |
0.0E+0 | -1.0E-3 | 1.0E-3 |
1.0E-1 | 9.0E-2 | 1.1E-1 |
1.0E+0 | 9.0E-1 | 1.1E+0 |
NIST Traceable Calibration
Test Point | Lower Tolerance | Upper Tolerance |
0.0E+0 | -1.0E-3 | 1.0E-3 |
1.0E-2 | 9.0E-3 | 1.1E-2 |
2.0E-2 | 1.8E-2 | 2.2E-2 |
5.0E-2 | 4.5E-2 | 5.5E-2 |
1.0E-1 | 9.0E-2 | 1.1E-1 |
2.0E-1 | 1.8E-1 | 2.2E-1 |
5.0E-1 | 4.5E-1 | 5.5E-1 |
1.0E+0 | 9.0E-1 | 1.1E+0 |
ISO 17025 Accredited Calibration
Test Point | Lower Tolerance | Upper Tolerance |
1.0E-3 | 0.0E-3 | 2.0E-3 |
1.0E-2 | 9.0E-3 | 1.1E-2 |
2.0E-2 | 1.8E-2 | 2.2E-2 |
5.0E-2 | 4.5E-2 | 5.5E-2 |
1.0E-1 | 9.0E-2 | 1.1E-1 |
2.0E-1 | 1.8E-1 | 2.2E-1 |
5.0E-1 | 4.5E-1 | 5.5E-1 |
1.0E+0 | 9.0E-1 | 1.1E+0 |
Section 5 – 4A Part 1
Applicable Controllers and Active Gauges
MX200, MX4A
Factory (Standard) Calibration
Test Point | Lower Tolerance | Upper Tolerance |
0.0E+0 | -1.0E-3 | 1.0E-3 |
1.0E+0 | 9.0E-1 | 1.1E+0 |
7.6E+2 | 6.8E+2 | 8.4E+2 |
NIST Traceable Calibration
Test Point | Lower Tolerance | Upper Tolerance |
0.0E+0 | -1.0E-3 | 1.0E-3 |
1.0E-2 | 9.0E-3 | 1.1E-2 |
5.0E-2 | 4.5E-2 | 5.5E-2 |
1.0E-1 | 9.0E-2 | 1.1E-1 |
5.0E-1 | 4.5E-1 | 5.5E-1 |
1.0E+0 | 9.0E-1 | 1.1E+0 |
1.0E+1 | 9.0E+0 | 1.1E+1 |
1.0E+2 | 9.0E+1 | 1.1E+2 |
7.6E+2 | 6.8E+2 | 8.4E+2 |
ISO 17025 Accredited Calibration
Test Point | Lower Tolerance | Upper Tolerance |
1.0E-3 | 0.0E-3 | 2.0E-3 |
1.0E-2 | 9.0E-3 | 1.1E-2 |
5.0E-2 | 4.5E-2 | 5.5E-2 |
1.0E-1 | 9.0E-2 | 1.1E-1 |
5.0E-1 | 4.5E-1 | 5.5E-1 |
1.0E+0 | 9.0E-1 | 1.1E+0 |
1.0E+1 | 9.0E+0 | 1.1E+1 |
1.0E+2 | 9.0E+1 | 1.1E+2 |
7.6E+2 | 6.8E+2 | 8.4E+2 |
Section 6 – 4A Part 2
Applicable Controllers and Active Gauges
MM200, MC300, MP4AR
Factory (Standard) Calibration
Test Point | Lower Tolerance | Upper Tolerance |
0.0E+0 | -1.0E-3 | 1.0E-3 |
1.0E+0 | 9.0E-1 | 1.1E+0 |
7.6E+2 | 6.8E+2 | 8.4E+2 |
NIST Traceable Calibration
Test Point | Lower Tolerance | Upper Tolerance |
0.0E+0 | -1.0E-3 | 1.0E-3 |
1.0E-2 | 9.0E-3 | 1.1E-2 |
5.0E-2 | 4.5E-2 | 5.5E-2 |
1.0E-1 | 9.0E-2 | 1.1E-1 |
5.0E-1 | 4.5E-1 | 5.5E-1 |
1.0E+0 | 9.0E-1 | 1.1E+0 |
1.0E+1 | 9.0E+0 | 1.1E+1 |
1.0E+2 | 8.0E+1 | 1.2E+2 |
7.6E+2 | 6.8E+2 | 8.4E+2 |
ISO 17025 Accredited Calibration
Test Point | Lower Tolerance | Upper Tolerance |
1.0E-3 | 0.0E-3 | 2.0E-3 |
1.0E-2 | 9.0E-3 | 1.1E-2 |
5.0E-2 | 4.5E-2 | 5.5E-2 |
1.0E-1 | 9.0E-2 | 1.1E-1 |
5.0E-1 | 4.5E-1 | 5.5E-1 |
1.0E+0 | 9.0E-1 | 1.1E+0 |
1.0E+1 | 9.0E+0 | 1.1E+1 |
1.0E+2 | 8.0E+1 | 1.2E+2 |
7.6E+2 | 6.8E+2 | 8.4E+2 |
Section 7 – 7B, 7E, 7M, 7F, 7FC, 7FCS Part 1
Applicable Controllers and Active Gauges
MX200, MM200, MC300, MP7ER, MP7FR
Factory (Standard) Calibration
Test Point | Lower Tolerance | Upper Tolerance |
1.0E-5 | 6.0E-6 | 1.6E-5 |
1.0E-4 | 6.0E-5 | 1.6E-4 |
NIST Traceable Calibration
Test Point | Lower Tolerance | Upper Tolerance |
1.0E-6 | 6.0E-7 | 1.6E-6 |
5.0E-6 | 3.2E-6 | 8.0E-6 |
1.0E-5 | 6.3E-6 | 1.6E-5 |
5.0E-5 | 3.2E-5 | 8.0E-5 |
1.0E-4 | 6.3E-5 | 1.6E-4 |
ISO 17025 Accredited Calibration
Test Point | Lower Tolerance | Upper Tolerance |
1.0E-6 | 6.0E-7 | 1.6E-6 |
5.0E-6 | 3.2E-6 | 8.0E-6 |
1.0E-5 | 6.3E-6 | 1.6E-5 |
5.0E-5 | 3.2E-5 | 8.0E-5 |
1.0E-4 | 6.3E-5 | 1.6E-4 |
Section 8 – 7B, 7E, 7M, 7F, 7FC, 7FCS Part 2
Applicable Controllers and Active Gauges
MX7B, MX7M
Factory (Standard) Calibration
Test Point | Lower Tolerance | Upper Tolerance |
1.0E-05 | 7.0E-06 | 1.3E-05 |
1.0E-04 | 7.0E-05 | 1.3E-04 |
NIST Traceable Calibration
Test Point | Lower Tolerance | Upper Tolerance |
1.0E-6 | 7.0E-7 | 1.3E-6 |
5.0E-6 | 3.5E-6 | 6.5E-6 |
1.0E-5 | 7.0E-6 | 1.3E-5 |
5.0E-5 | 3.5E-5 | 6.5E-5 |
1.0E-4 | 7.0E-5 | 1.3E-4 |
ISO 17025 Accredited Calibration
Test Point | Lower Tolerance | Upper Tolerance |
1.0E-6 | 7.0E-7 | 1.3E-6 |
5.0E-6 | 3.5E-6 | 6.5E-6 |
1.0E-5 | 7.0E-6 | 1.3E-5 |
5.0E-5 | 3.5E-5 | 6.5E-5 |
1.0E-4 | 7.0E-5 | 1.3E-4 |
Section 9 – CC-10
Applicable Controllers and Active Gauges
CC-10
Factory (Standard) Calibration
Test Point | Lower Tolerance | Upper Tolerance |
1.0E-5 | 6.0E-6 | 1.6E-5 |
1.0E+00 | 8.5E-1 | 1.2E+0 |
7.6E+2 | 7.2E+2 | 8.0E+2 |
NIST Traceable Calibration
Test Point | Lower Tolerance | Upper Tolerance |
1.0E-6 | 6.3E-7 | 1.6E-6 |
5.0E-6 | 3.2E-6 | 8.0E-6 |
1.0E-5 | 6.3E-6 | 1.6E-5 |
5.0E-5 | 3.2E-5 | 8.0E-5 |
1.0E-4 | 6.3E-5 | 1.6E-4 |
1.0E-3 | 6.3E-4 | 1.6E-3 |
1.0E-2 | 6.3E-3 | 1.6E-2 |
1.0E-1 | 8.5E-2 | 1.2E-1 |
1.0E+0 | 8.5E-1 | 1.2E+0 |
1.0E+1 | 8.5E+0 | 1.2E+1 |
1.0E+2 | 8.5E+1 | 1.2E+2 |
7.6E+2 | 7.2E+2 | 8.0E+2 |
ISO 17025 Accredited Calibration
Test Point | Lower Tolerance | Upper Tolerance |
1.0E-6 | 6.3E-7 | 1.6E-6 |
5.0E-6 | 3.2E-6 | 8.0E-6 |
1.0E-5 | 6.3E-6 | 1.6E-5 |
5.0E-5 | 3.2E-5 | 8.0E-5 |
1.0E-4 | 6.3E-5 | 1.6E-4 |
1.0E-3 | 6.3E-4 | 1.6E-3 |
1.0E-2 | 6.3E-3 | 1.6E-2 |
1.0E-1 | 8.5E-2 | 1.2E-1 |
1.0E+0 | 8.5E-1 | 1.2E+0 |
1.0E+1 | 8.5E+0 | 1.2E+1 |
1.0E+2 | 8.5E+1 | 1.2E+2 |
7.6E+2 | 7.2E+2 | 8.0E+2 |
Section 10 – 1E, 1F
Applicable Controllers and Active Gauges
MX200, MM200
Factory (Standard) Calibration
Test Point | Lower Tolerance | Upper Tolerance |
5.0E+0 | 4.0E+0 | 6.0E+0 |
1.0E+2 | 9.0E+1 | 1.1E+2 |
7.6E+2 | 7.2E+2 | 8.0E+2 |
NIST Traceable Calibration
Test Point | Lower Tolerance | Upper Tolerance |
5.0E+0 | 4.5E+0 | 5.5E+0 |
1.0E+2 | 9.0E+1 | 1.1E+1 |
4.0E+2 | 3.8E+2 | 4.2E+2 |
7.2E+2 | 7.2E+2 | 8.0E+2 |
ISO 17025 Accredited Calibration
Test Point | Lower Tolerance | Upper Tolerance |
5.0E+0 | 4.5E+0 | 5.5E+0 |
1.0E+2 | 9.0E+1 | 1.1E+2 |
4.0E+2 | 3.8E+2 | 4.2E+2 |
7.6E+2 | 7.2E+2 | 8.0E+2 |
Section 11 – CDG (Capacitance Diaphragm Gauge, Capacitance Manometer)
Applicable Controllers and Active Gauges
MX200, MM200
Note: FS indicates the full-scale range of the CDG (typically 0.1, 1, 10, 100, or 1000 Torr).
Factory (Standard) Calibration
Test Point | Lower Tolerance | Upper Tolerance |
0.1*FS | (0.1*FS) – 2% | (0.1*FS) + 2% |
0.5*FS | (0.5*FS) – 2% | (0.5*FS) + 2% |
FS | FS – 2% | FS + 2% |
NIST Traceable Calibration
Test Point | Lower Tolerance | Upper Tolerance |
0.01*FS | 0.01*FS – 0.001*FS | 0.01*FS + 0.001*FS |
0.1*FS | (0.1*FS) – 2% | (0.1*FS) + 2% |
0.2*FS | (0.2*FS) – 2% | (0.2*FS) + 2% |
0.3*FS | (0.3*FS) – 2% | (0.3*FS) + 2% |
0.4*FS | (0.4*FS) – 2% | (0.4*FS) + 2% |
0.5*FS | (0.5*FS) – 2% | (0.5*FS) + 2% |
0.6*FS | (0.6*FS) – 2% | (0.6*FS) + 2% |
0.7*FS | (0.7*FS) – 2% | (0.7*FS) + 2% |
0.8*FS | (0.8*FS) – 2% | (0.8*FS) + 2% |
0.9*FS | (0.9*FS) – 2% | (0.9*FS) + 2% |
FS | FS – 2% | FS + 2% |
ISO 17025 Accredited Calibration
Test Point | Lower Tolerance | Upper Tolerance |
0.01*FS | 0.01*FS – 0.001*FS | 0.01*FS + 0.001*FS |
0.1*FS | (0.1*FS) – 2% | (0.1*FS) + 2% |
0.2*FS | (0.2*FS) – 2% | (0.2*FS) + 2% |
0.3*FS | (0.3*FS) – 2% | (0.3*FS) + 2% |
0.4*FS | (0.4*FS) – 2% | (0.4*FS) + 2% |
0.5*FS | (0.5*FS) – 2% | (0.5*FS) + 2% |
0.6*FS | (0.6*FS) – 2% | (0.6*FS) + 2% |
0.7*FS | (0.7*FS) – 2% | (0.7*FS) + 2% |
0.8*FS | (0.8*FS) – 2% | (0.8*FS) + 2% |
0.9*FS | (0.9*FS) – 2% | (0.9*FS) + 2% |
FS | FS – 2% | FS + 2% |