This application note explains the process for cleaning the Televac® 7B cold cathode vacuum gauge and 7E cold cathode vacuum gauge.

Istruzioni

1. Rimuovere il sensore dalla camera a vuoto.

2. Unscrew the cap to expose the anode assembly (to remove, rotate the cap counter clockwise).

3. Sollevare delicatamente il gruppo anodico verso l'alto per separarlo dal corpo.

4. Remove the “O” ring from the body, being careful not to scratch the “O” ring seat. Wipe it off with a lint-free tissue making sure that it is still intact (no cracks or tears). It should be round and not flattened.

5. Blast the entire anode assembly at 30 psi using 70-140 mesh glass bead abrasive, being careful not to damage the glass feedthrough. (If any rainbow effect is noticed on the glass bead, a crack in the feedthrough may exist and the anode assembly should be replaced (PN 2-6200-37 for the 7B and PN 2-7900-51 for the 7E).

6. Blast the body with glass beads concentrating on the pole piece and “O” ring areas. Clean the remaining internal parts. If the threaded end of the tube needs cleaning, it can be bead blasted as well.

7. After cleaning, blow out any residual glass beads or dust with dry air. See that all metal particles are removed from the pole piece areas.

8. Re-grease the “O” ring with Apiezon L or M Grease (a vacuum approved grease). DO NOT APPLY EXCESS GREASE, “O” ring should be shiny.

9. Install “O” ring in the body.

10. Replace the anode assembly in the body after the “O” ring has been installed. Observe the keyway location.

11. Screw on the cap and hand tighten.

12. Reinstallare nella camera a vuoto.

13. Allow several hours for the tube to degas when pumped down to high vacuum.

14. Se pulito e assemblato correttamente, il tubo del misuratore è pronto per l'uso senza ricalibrazione, a meno che non sia necessaria una calibrazione tracciabile NIST.
a meno che non sia necessaria una calibrazione tracciabile NIST.

Guardate il video dimostrativo

Prodotti correlati

Misuratore di vuoto a catodo freddo NW25/KF25 in acciaio inox 7B Penning Magnetron - Numero di parte: 2-2100-272

Misuratore di vuoto a catodo freddo a catodo freddo con magnete a penna 7B

- 1*10-7 Torr a 1*10-3 Torr
- Robusto e altamente resistente alla contaminazione
- Facile da smontare e pulire

Vacuometro a doppio magnetron invertito 7E

Vacuometro a catodo freddo 7E

• 1*10-8 Torr to 1*10-2 Torr
• Contamination resistant, resistant to inrushes of gas
• Easy to disassemble and clean