Descrizione
Historically the 2A thermocouple sensor had a measurement range up to only 20 Torr and the 4A convection sensor was used to measure pressures up to 1000 Torr. Software and drive signal enhancements incorporated into the MX2A active thermocouple gauge now allow the 2A thermocouple sensor to measure pressures up to 1000 Torr. This has raised questions about the differences between the MX2A thermocouple active gauge and the MX4A active convection active gauge as almost all of their specifications are now identical. This document outlines the differences between the two active gauges.
Differenze tra MX2A e MX4A
Descrizione | MX2A | MX4A |
Tipo di sensore | Termocoppia | Convezione |
Diametro del filamento | Più spessa | Diluente |
Resistenza alla contaminazione | Più in alto | Abbassare |
Tempo di risposta | <2 seconds | <1 second |
Precisione | ±30% (da 10 a 100 Torr) | ±10% (da 10 a 100 Torr) |
Materiali esposti al vuoto | Lega NiMnSi, lega di CuNiMn | Lega NiCr, Lega CuNi |
Costo | Il più basso | Basso |
Prodotti correlati
Misuratore di vuoto attivo a termocoppia MX2A
- 1*10-4 Torr fino a 1000 Torr
- Display OLED a colori di facile lettura
- Sensori facili da sostituire
Misuratore di vuoto attivo a convezione MX4A
- 1*10-4 Torr a 1000 Torr
- Sensori facili da sostituire
- Resistente alla contaminazione