Televac AN 3001: Confronto tra i manometri attivi MX2A e MX4A

Descrizione

Historically the 2A thermocouple sensor had a measurement range up to only 20 Torr and the 4A convection sensor was used to measure pressures up to 1000 Torr. Software and drive signal enhancements incorporated into the MX2A active thermocouple gauge now allow the 2A thermocouple sensor to measure pressures up to 1000 Torr. This has raised questions about the differences between the MX2A thermocouple active gauge and the MX4A active convection active gauge as almost all of their specifications are now identical. This document outlines the differences between the two active gauges.

Differenze tra MX2A e MX4A

Descrizione MX2A MX4A
Tipo di sensore Termocoppia Convezione
Diametro del filamento Più spessa Diluente
Resistenza alla contaminazione Più in alto Abbassare
Tempo di risposta <2 seconds <1 second
Precisione ±30% (da 10 a 100 Torr) ±10% (da 10 a 100 Torr)
Materiali esposti al vuoto Lega NiMnSi, lega di CuNiMn Lega NiCr, Lega CuNi
Costo Il più basso Basso

Tutte le altre specifiche per MX2A e MX4A sono equivalenti.

Prodotti correlati

Misuratore di vuoto attivo a termocoppia MX2A

Misuratore di vuoto attivo a termocoppia MX2A

- 1*10-4 Torr fino a 1000 Torr
- Display OLED a colori di facile lettura
- Sensori facili da sostituire

Vacuometro digitale attivo a convezione MX4A

Misuratore di vuoto attivo a convezione MX4A

- 1*10-4 Torr a 1000 Torr
- Sensori facili da sostituire