Televac AN 3006: Azzeramento della memoria dati MM200

Descrizione

This application note explains the process for resetting the MM200 data memory from a Windows PC. The serial communications program Tera Term must be installed prior to following these instructions, and a serial to USB converter must be purchased. Note that this will reset all configurable features of the MM200, such as analog output and set point settings. If you don’t have Tera Term installed, click here to install the latest version from GitHub.

Istruzioni

1. Plug the serial to USB converter into your PC.

2. Aprire il menu Start.

3. Aprire il Pannello di controllo.

4. Aprite Hardware e suoni.

5. Aprire la Gestione periferiche.

6. Espandere il menu a tendina Porte (COM e LPT).

7. Determine the COM port assignment for your USB to serial converter, in this example the converter is assigned to COM3.

8. Open TeraTerm.

9. Aprire il menu Impostazione e selezionare l'opzione Porta seriale....

10. Change the Port setting to the COM port for your USB to serial converter, in this example it is COM3.

11. Ensure that the remainder of the settings are the following, then select OK:

  1. Velocità di trasmissione: 9600
  2. Dati: 8 bit
  3. Parità: nessuna
  4. Stop: 1 bit
  5. Controllo di flusso: nessuno
  6. Ritardo di trasmissione: 0 msec/char, 0 msec/linea

12. Open the Setup menu and select the Terminal setup… option.

13. Assicurarsi che le impostazioni siano le seguenti, quindi selezionare OK:

  1. Ricevere: CR+LF
  2. Trasmettere: CR+LF
  3. Controllo dell'eco locale

14. Send the following commands (<CR> is equivalent to pressing the Enter key):

  1. D1<CR>
  2. CF<CR>
  3. PF<CR>
  4. BF<CR>
  5. CA3<CR>
  6. SE<CR>

15. The memory has been reset and the MM200 should be functional. If you are still experiencing issues, start a online chat with us, or contact our Technical Support Team.