Televac Tolerances

Accuracy Specification and Vacuum Test Points with Tolerances

Description

All values are in Torr unless otherwise stated and FS denotes the full scale or maximum reading of the CDG. Televac reserves the right to change tolerances for any controllers, active gauges, or sensors. Any modification to any unit without approval from Televac® voids the unit’s certificate of compliance. Additional NIST Calibration Test Points are available upon request.

Applicable Gauges

  • MC300 2A, 2A VacuMini, or 2A NASA Mini
  • MM200 2A, 2A VacuMini, or 2A NASA Mini
  • MX200 2A, 2A VacuMini, or 2A NASA Mini
  • MV2A, 2A VacuMini, or 2A NASA Mini
  • Compact 2A, 2A VacuMini, or 2A NASA Mini
  • Vacuguard™, 2A VacuMini, or 2A NASA Mini
  • B2A Portable, 2A VacuMini, or 2A NASA Mini
  • MP2AR

Accuracy Specification

Range Tolerance
1.0E-3 to 1.0E-2 ±1.0E-3
1.0E-2 to 1.0E-1 ±10%
1.0E-1 to 1.0E+0 ±20%

Standard Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
0.0E+0 -1.0E-3 1.0E-3
1.0E-1 9.0E-2 1.1E-1
1.0E+0 8.0E-1 1.2E+0

NIST Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
0.0E+0 -1.0E-3 1.0E-3
1.0E-2 9.0E-3 1.1E-2
2.0E-2 1.8E-2 2.2E-2
5.0E-2 4.5E-2 5.5E-2
1.0E-1 9.0E-2 1.1E-1
2.0E-1 1.6E-1 2.4E-1
5.0E-1 4.0E-1 6.0E-1
1.0E+0 8.0E-1 1.2E+0

GE Aviation Specification P10TF3 Compatible Calibration

The 2A sensor in combination with any of the controllers above can be calibrated to the GE Aviation Specification P10TF3.

Test Point Lower Tolerance Upper Tolerance
0.0E+0 -1.0E-3 1.0E-3
1.0E-2 9.0E-3 1.1E-2
2.0E-2 1.8E-2 2.2E-2
5.0E-2 4.5E-2 5.5E-2
1.0E-1 9.0E-2 1.1E-1
2.0E-1 1.7E-1 2.3E-1
5.0E-1 4.3E-1 5.8E-1
1.0E+0 8.5E-1 1.2E+0

Applicable Gauges

  • MX2A

Accuracy Specification

Range Tolerance
1.0E-3 to 1.0E-2 ±1.0E-3
1.0E-2 to 1.0E+1 ±10%
1.0E+0 to 1.0E+2 ±30%
1.0E+2 to 1.0E+3 ±10%

Standard Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
0.0E+0 -1.0E-3 1.0E-3
1.0E-1 9.0E-2 1.1E-1
1.0E+0 9.0E-1 1.1E+0

NIST Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
0.0E+0 -1.0E-3 1.0E-3
1.0E-2 9.0E-3 1.1E-2
2.0E-2 1.8E-2 2.2E-2
5.0E-2 4.5E-2 5.5E-2
1.0E-1 9.0E-2 1.1E-1
2.0E-1 1.8E-1 2.2E-1
5.0E-1 4.5E-1 5.5E-1
1.0E+0 9.0E-1 1.1E+0

Applicable Gauges

  • MC300 4A
  • MM200 4A
  • MP4AR

Accuracy Specification

Range Tolerance
1.0E-3 to 1.0E-2 ±1.0E-3
1.0E-2 to 1.0E+1 ±10%
1.0E+1 to 1.0E+2 ±20%
1.0E+2 to 1.0E+3 ±10%

Standard Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
0.0E+0 -1.0E-3 1.0E-3
1.0E+0 9.0E-1 1.1E-0
7.6E+2 6.8E+2 8.4E+2

NIST Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
0.0E+0 -1.0E-3 1.0E-3
1.0E-2 9.0E-3 1.1E-2
5.0E-2 4.5E-2 5.5E-2
1.0E-1 9.0E-2 1.1E-1
5.0E-1 4.5E-1 5.5E-1
1.0E+0 9.0E-1 1.1E+0
1.0E+1 9.0E+0 1.1E+1
1.0E+2 8.0E+1 1.2E+2
7.6E+2 6.8E+2 8.4E+2

Applicable Gauges

  • MX4A
  • MX200 4A

Accuracy Specification

Range Tolerance
1.0E-3 to 1.0E-2 ±1.0E-3
1.0E-2 to 1.0E+3 ±10%

Standard Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
0.0E+0 -1.0E-3 1.0E-3
1.0E+0 9.0E-1 1.1E+0
7.6E+2 6.8E+2 8.4E+2

NIST Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
0.0E+0 -1.0E-3 1.0E-3
1.0E-2 9.0E-3 1.1E-2
5.0E-2 4.5E-2 5.5E-2
1.0E-1 9.0E-2 1.1E-1
5.0E-1 4.5E-1 5.5E-1
1.0E+0 9.0E-1 1.1E+0
1.0E+1 9.0E+0 1.1E+1
1.0E+2 9.0E+1 1.1E+2
7.6E+2 6.8E+2 8.4E+2

Applicable Gauges

  • MC300 7B
  • MM200 7B
  • MX200 7B

Accuracy Specification

Range Tolerance
1.0E-6 to 1.0E-3 ±.2 decade (-37% to 58%)

Standard Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
1.0E-5 6.0E-6 1.6E-5
1.0E-4 6.0E-5 1.6E-4

NIST Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
1.0E-6 6.3E-7* 1.6E-6
5.0E-6 3.2E-6 8.0E-6
1.0E-5 6.3E-6 1.6E-5
5.0E-5 3.2E-5 8.0E-5
1.0E-4 6.3E-5 1.6E-4

*The tolerance for the MM200 and MC300 is 6E-7


Applicable Gauges

  • MC300 7E
  • MM200 7E
  • MX200 7E
  • MP7ER

Accuracy Specification

Range Tolerance
1.0E-7 to 1.0E-2 ±.2 decade (-37% to 58%)

Standard Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
1.0E-5 6.3E-6 1.6E-5
1.0E-4 6.3E-5 1.6E-4

NIST Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
1.0E-6 6.3E-7 1.6E-6
5.0E-6 3.2E-6 8.0E-6
1.0E-5 6.3E-6 1.6E-5
5.0E-5 3.2E-5 8.0E-5
1.0E-4 6.3E-5 1.6E-4

Applicable Gauges

  • MC300 7F/7FC
  • MM200 7F/7FC/7FCS
  • MX200 7F/7FC/7FCS
  • MP7FR

Accuracy Specification

Range Tolerance
1.0E-9 to 1.0E-2 ±.2 decade (-37% to 58%)

Standard Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
1.0E-5 6.3E-6 1.6E-5
1.0E-4 6.3E-5 1.6E-4

NIST Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
1.0E-6 6.3E-7 1.6E-6
5.0E-6 3.2E-6 8.0E-6
1.0E-5 6.3E-6 1.6E-5
5.0E-5 3.2E-5 8.0E-5
1.0E-4 6.3E-5 1.6E-4

Applicable Gauges

  • CC-10

Accuracy Specification

Range Tolerance
1.0E-9 to 1.0E-1 ±.2 decade (-37% to 58%)
1.0E-1 to 2.0E+2 ±15%
2.0E+2 to 1.0E+3 ±5%

Standard Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
1.0E-05 6.0E-6 1.6E-5
1.0E+00 8.5E-1 1.2E+0
7.6E+02 7.2E+2 8.0E+2

NIST Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
1.0E-6 6.3E-7 1.6E-6
5.0E-6 3.2E-6 8.0E-6
1.0E-5 6.3E-6 1.6E-5
5.0E-5 3.2E-5 8.0E-5
1.0E-4 6.3E-5 1.6E-4
1.0E-3 6.3E-4 1.6E-3
1.0E-2 6.3E-3 1.6E-2
1.0E-1 8.5E-2 1.2E-1
1.0E+0 8.5E-1 1.2E+0
1.0E+1 8.5E+0 1.2E+1
1.0E+2 8.5E+1 1.2E+2
7.6E+2 7.2E+2 8.0E+2

Applicable Gauges

  • MM200 1E
  • MX200 1E

Accuracy Specification

Range Tolerance
1.0E+0 to 1.0E+1 ±1.0E+0
1.0E+1 to 1.0E+2 ±10%
1.0E+2 to 1.0E+3 ±5%

Standard Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
5.0E+00 4.0E+0 6.0E+0
1.0E+2 9.0E+1 1.1E+2
7.6E+2 7.2E+2 8.0E+2

NIST Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
5.0E+00 4.5E+0 5.5E+0
1.0E+2 9.0E+1 1.1E+2
4.0E+2 3.8E+2 4.2E+2
7.6E+2 7.2E+2 8.0E+2

Applicable Gauges

  • MP3DR
  • MC300 Mini BA
  • MM200 Mini BA

Accuracy Specification

Range Tolerance
1.0E-9 to 1.0E-2 ±.1 decade (-21% to 26%)

Standard Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
1.0E-05 7.9E-06 1.3E-05
1.0E-04 7.9E-05 1.3E-04

NIST Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
1.0E-06 7.9E-07 1.3E-06
5.0E-06 4.0E-06 6.3E-06
1.0E-05 7.9E-06 1.3E-05
5.0E-05 4.0E-05 6.3E-05
1.0E-04 7.9E-05 1.3E-04

Applicable Gauges

MM200 Capacitive Diaphragm Gauge
MX200 Capacitive Diaphragm Gauge

Accuracy Specification

Range Tolerance
.01*FS to .1*FS ±.001*FS
.1*FS to FS ±2%

Standard Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
0.1*FS (0.1*FS)-2% (0.1*FS)+2%
0.5*FS (0.5*FS)-2% (0.5*FS)+2%
FS FS-2% FS+2%

NIST Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
0.01*FS 0.01*FS-.001*FS 0.01*FS +.001*FS
0.1*FS (0.1*FS)-2% (0.1*FS)+2%
0.2*FS (0.2*FS)-2% (0.2*FS)+2%
0.3*FS (0.3*FS)-2% (0.3*FS)+2%
0.4*FS (0.4*FS)-2% (0.4*FS)+2%
0.5*FS (0.5*FS)-2% (0.5*FS)+2%
0.6*FS (0.6*FS)-2% (0.6*FS)+2%
0.7*FS (0.7*FS)-2% (0.7*FS)+2%
0.8*FS (0.8*FS)-2% (0.8*FS)+2%
0.9*FS (0.9*FS)-2% (0.9*FS)+2%
FS FS-2% FS+2%

Applicable Gauges

  • MX7B
  • MX7M

Accuracy Specification

Range Tolerance
1.0E-7 to 1.0E-3 ±30%

Standard Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
1.0E-05 7.0E-06 1.3E-05
1.0E-04 7.0E-05 1.3E-04

NIST Calibration Points with Tolerances

Test Point Lower Tolerance Upper Tolerance
1.0E-06 7.0E-07 1.3E-06
5.0E-06 3.5E-06 6.5E-06
1.0E-05 7.0E-06 1.3E-05
5.0E-05 3.5E-05 6.5E-05
1.0E-04 7.0E-05 1.3E-04