Televac AN 3011: MX200 및 MM200 아날로그 출력 비교

설명

This application note shows the MX200 Vacuum Controller analog (0 to 10 V DC) output settings that are equivalent to an MM200 Vacuum Controller.

비교 차트

녹색 강조 표시값은 해당 센서 유형에 대한 기본 MX200 설정입니다.

빨간색 강조 표시값은 해당 센서 유형에 대한 기본 MM200 설정입니다.

센서 M200 Code MX200 설정 높은 가치 낮은 값
1E R1EN Linear (LIN) 1.0E+03 Torr 1.0E+00 Torr
R1EL Not supported
1F R1EN Linear (LIN) 1.0E+04 Torr 1.0E+01 Torr
R1EL Not supported
2A R2AN Linear (LIN) 1.0E+00 Torr 1.0E-03 Torr
R2AL Logarithmic (LOG) 1.0E+00 Torr 1.0E-03 Torr
R2AT Not supported
R2AC Not supported
R2AR Not supported
4A R4AN Linear-by-decade (L-D) 1.0E+03 Torr 1.0E-03 Torr
R4AU Linear (LIN) 1.0E+00 Torr 1.0E-03 Torr
R4AT Not supported
R4AL Logarithmic (LOG) 1.0E+03 Torr 1.0E-03 Torr
5A/5B/5C/5D/5E/5F R5AN Linear (LIN) Full scale Full scale * 0.001
R5AL Not supported
7B R7BM Linear-by-decade (L-D) 1.0E-03 1.0E-07
R7BL Logarithmic (LOG) 1.0E-03 1.0E-07
R7BE Not supported
R7BN Linear (LIN) 1.0E-03 1.0E-06
R7BA Not supported
R7BT Not supported
7E R7EM Linear-by-decade 7E (L-D 7E) 1.0E-02 1.0E-08
R7EL Logarithmic (LOG) 1.0E-02 1.0E-11
R7Ex Linear (LIN) 1.0E-0x* 1.0E-(x+3)*
7F/7FC/7FCS R7FM Linear-by-decade (L-D) 1.0E-02 1.0E-11
R7FL Logarithmic (LOG) 1.0E-02 1.0E-11
R7Fx Linear (LIN) 1.0E-0x* 1.0E-(x+3)*

*x값에 따라 달라집니다. 7E = 2≤x≤5, 7F/7FC/7FCS = 2≤x≤8.

관련 제품

1E-11 ~ 10,000 토르의 전체 진공 범위를 제공하는 모듈식 진공 컨트롤러

MX200 진공 컨트롤러

• 1*10-11 Torr  to 1*104 Torr
• RS-232/RS-485/USB digital communications
• Control up to 10 Televac® vacuum gauges
• Easy-to-read OLED display

Televac®️ MM200 진공 컨트롤러

MM200 진공 컨트롤러

• 고객 구성 가능한 설계
• RS-232/RS-485 디지털 통신
• 최대 8개의 릴레이 설정점 지원
• 선택 가능한 측정 단위