Televac® Compact 2A Thermocouple (Pirani) Vacuum Controller

Part Number: 2-3014-042


Compatible With: 2A Thermocouple, 2A VacuMini Thermocouple, 2A NASA Mini Thermocouple


Features and Benefits:

  • Panel mountable
  • Analog 0 to 10 V DC output for simple integration
  • Analog sweeping needle display
  • Excellent customer support
  • NIST traceable calibration optional

The Televac® Compact 2A vacuum controller (also called a vacuum control unit or vacuum pressure controller) can be connected to 1 Televac® passive rough vacuum gauge, like the 2A thermocouple2A VacuMini Thermocouple, or 2A NASA Mini Thermocouple. It includes a 0 to 10 V DC analog output the connected vacuum gauge for simple integration with a PC or PLC, and acts as a vacuum pressure controller for process control in your vacuum system.

The Compact 2A vacuum controller’s user interface includes an analog sweeping needle display, which shows measurements from 1 mTorr (1 micron) up to 20 Torr. This vacuum control unit supports one gas type of nitrogen/air. The Compact 2A vacuum controller is a versatile unit, making it suitable for various vacuum measurement systems specific to your needs.

Looking for a more modular, digital version of this vacuum controller? Check out the MX200. Need a single-channel vacuum controller with a digital display? Check out the MV2A.

ImageCompact 2A ThermocouplePart NumberPriceBuy
Compact 2A Analog Vacuum Controller
Compact 2A Thermocouple2-3014-042$540.00

2A Cable

ImageLengthPart NumberPriceBuy
Televac® VacuGuard™ 2A Modular Connector
10'2-9800-077$130.00
Televac® VacuGuard™ 2A Modular Connector
20'2-9800-078$155.00
Televac® VacuGuard™ 2A Modular Connector
35'2-9800-079$190.00
Televac® VacuGuard™ 2A Modular Connector
50'2-9800-080$225.00
Televac® VacuGuard™ 2A 10' Coiled Modular Connector
10' Coiled2-9800-081$130.00

2A Mini Cable

ImageLengthPart NumberPriceBuy
2A Mini Modular Connector (MC300, MV2A, Compact 2A) - 10'
10'2-9800-096$205.00
2A Mini Coiled Modular Connector
10' Coiled2-9800-082$190.00
Operating Range (Torr) 1*10-3 to 20 Torr
Communications None
Analog Output 0 V to 10 V DC
Supply Voltage 110/220 V AC, 50-60 Hz
Calibration Medium Dry air or nitrogen
Operating Temperature 0 °C to 50 °C
Storage Temperature -40 °C to 85 °C
Display Readable Distance 3 m (10 ft)
Accuracy
1 mTorr to 10 mTorr ±1 mTorr
10 mTorr to 100 mTorr ±10% of reading
100 mTorr to 1 Torr ±20% of reading
1 Torr to 20 Torr Indication only
Analog Output ±5 mV