This application note explains the process for cleaning the Televac® 7FC cold cathode and 7FCS cold cathode vacuum sensors that have been used in harsh environments or operated for extended periods in the 10-4 Torr or higher range.

Необходимые материалы

Для правильной очистки датчиков холодного катода 7FC и вакуумных датчиков холодного катода 7FCS необходимы следующие материалы:

1. 3/16” Allen Wrench

Инструкции

1. Remove the sensor from the vacuum chamber.

2. Unscrew the bolts to expose the anode assembly.

3. Gently lift the anode assembly upwards to separate it from the body.

4. Remove the copper gasket from the sensor.

5. Check the knife-edge on both the anode and the body to make sure they are in good condition.

6. Gently remove any contaminants from the anode using the scouring pad. For the 7FCS, be careful not to damage the starting aid.

7. Blast the body with glass beads.

8. After cleaning, blow out any residual glass beads or dust with dry nitrogen.

9. Install a new copper gasket in body.

10. Replace the anode assembly in the body.

11. Tighten bolts gradually moving from bolt to bolt to ensure uniform compression of the gasket.

12. Reinstall in the vacuum chamber.

13. Allow several hours for tube to degas when pumped down to high vacuum.

14. If properly cleaned and assembled, the gauge tube is ready for use without re-calibration unless NIST
traceable calibration is required.

Watch The Video

Похожие продукты

7FCS DI Магнетронный вакуумметр с холодным катодом

7FC Cleanable Cold Cathode Vacuum Gauge
• 1*10-11 Torr to 1*10-2 Torr
• Cleanable for extended sensor life
• Resistant to inrushes of gas

7FCS DI Магнетронный вакуумметр с холодным катодом

7FCS Быстрый старт Вакуумметр с холодным катодом

- 1*10-11 Торр до 1*10-2 Торр
- Быстрое зажигание при СВН
- Устойчивый к выбросам газа
- Очищается для продления срока службы датчика