Televac AN 3002:清洁7B和7E冷阴极无源仪表

说明

This application note explains the process for cleaning the Televac® 7B cold cathode vacuum gauge and 7E cold cathode vacuum gauge.

说明

1.从真空室中取出传感器。

2. Unscrew the cap to expose the anode assembly (to remove, rotate the cap counter clockwise).

3.轻轻向上提起阳极组件,使其与机体分离。

4. Remove the “O” ring from the body, being careful not to scratch the “O” ring seat. Wipe it off with a lint-free tissue making sure that it is still intact (no cracks or tears). It should be round and not flattened.

5. Blast the entire anode assembly at 30 psi using 70-140 mesh glass bead abrasive, being careful not to damage the glass feedthrough. (If any rainbow effect is noticed on the glass bead, a crack in the feedthrough may exist and the anode assembly should be replaced (PN 2-6200-37 for the 7B and PN 2-7900-51 for the 7E).

6. Blast the body with glass beads concentrating on the pole piece and “O” ring areas. Clean the remaining internal parts. If the threaded end of the tube needs cleaning, it can be bead blasted as well.

7. After cleaning, blow out any residual glass beads or dust with dry air. See that all metal particles are removed from the pole piece areas.

8. Re-grease the “O” ring with Apiezon L or M Grease (a vacuum approved grease). DO NOT APPLY EXCESS GREASE, “O” ring should be shiny.

9. Install “O” ring in the body.

10. Replace the anode assembly in the body after the “O” ring has been installed. Observe the keyway location.

11. Screw on the cap and hand tighten.

12.重新安装在真空室中。

13. Allow several hours for the tube to degas when pumped down to high vacuum.

14.如果清洗和装配正确,测量管就可以直接使用,无需重新校准,除非需要NIST
可追溯的校准。