Televac AN 3022: 用MX2A/MX4A自动控制MP7ER/MP7FR。

宗旨

The application note explains how the open collector set point of an MX2A or MX4A rough vacuum active gauge can be used to automatically control the turn on/off of an MP7ER or MP7FR cold cathode high vacuum active gauge.

Relevant Televac Products

This document applies to the Televac® products listed

说明 部件号
MX2A 2-8910-1XX
MX4A 2-8930-1XX
MP7ER 2-7950-XXX
MP7FR 2-7970-XXX
MX2A/MX4A 6导体电缆 2-9857-XXX
MP7ER/MP7FR 9导体电缆。 2-9852-XXX

背景介绍

The Televac® MX2A and MX4A are thermal conductivity active gauges for rough vacuum measurement (1*10-3 Torr to 1000 Torr) and the Televac® MP7ER and MP7FR are cold cathode active gauges for high vacuum measurement (1*10-11/1*10-8 Torr to 1*10-2 Torr).

由于冷阴极的设计,在1*10-2Torr以上的压力下不开启冷阴极是很重要的。这有助于防止溅射,以及在较高压力下开启冷阴极时发生的氧化和污染,从而导致读数不准确和损坏仪表。

Because the cold cathode can’t read above 1*10-2 Torr, a rough vacuum gauge such as the MX2A or MX4A must be used to switch the cold cathode on/off at 1*10-2 Torr.

The MX2A and MX4A both include an open collector set point, which is essentially a switch that can be set to a specific vacuum reading. In this case, our set point is set to 1*10-2 Torr. This set point switch can be wired directly to the cold cathode turn on/off control pin of the MP7ER or MP7FR, creating a standalone active vacuum gauge system capable of measuring from 1*10-11/1*10-8 Torr to 1000 Torr.

Setting up the MX2A/MX4A Set Point 1 (Open Collector)

MX2A/MX4A设定点1开式集电极有两种设置。

1.SP1H(设定点1高)--设定点停用的真空读数。
2.SP1L(设定点1低)--设定点激活的真空读数。

当设定点用于控制MP7ER/MP7FR时,我们推荐SP1H和SP1L的设置如下。

1.SP1H-0.012托尔(相当于12毫托尔、12微米和1.2*10-2托尔)
2.SP1L-0.010托尔(相当于10毫托尔、10微米和1.0*10-2托尔)

请注意,这些是建议,您的系统可能需要不同的设置。
要更改SP1H和SP1L设置,请按照以下步骤进行。

1.在测量屏幕上,按两次SEL键。
2.2. 屏幕将显示"Setpoints SP1L .XXX"(其中.XXX为当前设置)。
3.按ENT键,屏幕上方中央将显示"ADJ"。
4.使用向上和向下箭头调整数值,直到读出".010"。
5.按ENT键,屏幕上方的"ADJ"将随之消失。
6.按向下箭头,屏幕将显示"Setpoints SP1H .XXX"(其中.XXX为当前设置)。
7.按ENT键,屏幕上方中央再次显示"ADJ"。
8.使用向上和向下箭头调整数值,直到读出".012"。
9.按ENT键,屏幕上方的"ADJ"将随之消失。
10.按SEL键三次,返回测量屏幕(或等待超时)。

The SP1H and SP1L settings can also be changed via RS-485 communications using the W2 command. The command to change the SP1H and SP1L settings to the values recommended above is the following (where X is the address of the unit, default is 0, and <cr> is a carriage return which is equivalent to pressing the enter key on your keyboard):

1. *XW210021202<cr>

Connecting the MX2A/MX4A and MP7ER/MP7FR

These instructions assume the use of Televac® MX2A/MX4A cables PN 2-9857-XXX and the Televac® MP7ER/MP7FR cables PN 2-9852-XXX. Once you’ve completed the MX2A/MX4A setup described in Section 3, make the connections listed in the table below. What this setup will do is supply power to both units, ground the source connection of the MX2A/MX4A set point 1, and connect the open drain of the MX2A/MX4A set point 1 to the cold cathode turn on/off control pin of the MP7ER/MP7FR.

当系统被抽空并达到1.0*10-2Torr时,MX2A/MX4A的设定点1将切换,将MP7ER/MP7FR的冷阴极接通/断开控制引脚与地连接,使其接通。反之,当系统通风或漏风,达到1.2*10-2Torr时,控制引脚将与地断开,关闭冷阴极。

注:

1.图中所示的导线之间不需要直接连接,例如可以通过电子柜中的接线端子进行接线。
2.2. 系统可能需要额外的连接;例如,模拟0至10 V直流输出线可能需要连接到PLC的模拟输入,或者RS-485线可能需要连接到PLC。

MX2A/MX4A MP7ER/MP7FR 说明
引脚5(绿色 引脚1(棕色 设定点1连接到冷阴极开/关控制引脚。
引脚12(黑色 查看说明 接电源(-)地线,任意连接。
引脚8(红色 引脚4(黄色 电源 (+) 24 V DC, 这些可以独立连接。
引脚15(黑色 引脚2(红色 电源(-)地,可独立连接。

Additional Considerations

这些说明描述了MX2A/MX4A可以用来自动切换MP7ER/MP7FR的冷阴极控制引脚的一种方法。还有其他各种可能更适合您独特的真空系统的方法,包括以下几种。

1.用MX2A/MX4A设定点2继电器切换MP7ER/MP7FR的冷阴极控制脚。
2.从PLC(或其他控制系统)监测MX2A/MX4A的模拟输出,将模拟输出转换为真空读数,用PLC直接切换MP7ER/MP7FR的冷阴极控制引脚的1*10-2托(10微米)。
3.与方案2类似,但用RS-485通讯方式监测MX2A/MX4A的真空读数。

相关产品

MX2A 热电偶主动真空计

MX2A 热电偶主动真空计

- 1*10-4托尔至1000托尔。
- 易于阅读的彩色OLED显示屏
- 传感器易于更换

MX4A对流式主动数字真空计

MX4A对流主动真空计

- 1*10-4托尔至1000托尔。
- 传感器易于更换
- 抗污染

MP7ER有源冷阴极真空计

MP7ER Cold Cathode Active Gauge

• 1*10-8 Torr to 1*10-2 Torr
• Cleanable sensor for extended use
• Minimal stray magnetic field

MP7ER有源冷阴极真空计

MP7FR Cold Cathode Active Gauge

• 1*10-11 Torr to 1*10-2 Torr
• Compact and robust design
• Minimal stray magnetic field