Televac AN 3011: MX200 and MM200 Analog Output Comparisons

Description

This application note shows the MX200 Vacuum Controller analog (0 to 10 V DC) output settings that are equivalent to an MM200 Vacuum Controller.

Comparison Chart

Green highlighted values are the default MX200 settings for that sensor type

Red highlighted values are the default MM200 settings for that sensor type

Sensor M200 Code MX200 Setting High Value Low Value
1E R1EN Linear (LIN) 1.0E+03 Torr 1.0E+00 Torr
R1EL Not supported
1F R1EN Linear (LIN) 1.0E+04 Torr 1.0E+01 Torr
R1EL Not supported
2A R2AN Linear (LIN) 1.0E+00 Torr 1.0E-03 Torr
R2AL Logarithmic (LOG) 1.0E+00 Torr 1.0E-03 Torr
R2AT Not supported
R2AC Not supported
R2AR Not supported
4A R4AN Linear-by-decade (L-D) 1.0E+03 Torr 1.0E-03 Torr
R4AU Linear (LIN) 1.0E+00 Torr 1.0E-03 Torr
R4AT Not supported
R4AL Logarithmic (LOG) 1.0E+03 Torr 1.0E-03 Torr
5A/5B/5C/5D/5E/5F R5AN Linear (LIN) Full scale Full scale * 0.001
R5AL Not supported
7B R7BM Linear-by-decade (L-D) 1.0E-03 1.0E-07
R7BL Logarithmic (LOG) 1.0E-03 1.0E-07
R7BE Not supported
R7BN Linear (LIN) 1.0E-03 1.0E-06
R7BA Not supported
R7BT Not supported
7E R7EM Linear-by-decade 7E (L-D 7E) 1.0E-02 1.0E-08
R7EL Logarithmic (LOG) 1.0E-02 1.0E-11
R7Ex Linear (LIN) 1.0E-0x* 1.0E-(x+3)*
7F/7FC/7FCS R7FM Linear-by-decade (L-D) 1.0E-02 1.0E-11
R7FL Logarithmic (LOG) 1.0E-02 1.0E-11
R7Fx Linear (LIN) 1.0E-0x* 1.0E-(x+3)*

*Dependent on the value of x; 7E = 2≤x≤5, 7F/7FC/7FCS = 2≤x≤8.

Related Products

Modular vacuum controller that offers the full vacuum range of 1E-11 to 10,000 Torr

MX200 Vacuum Controller

• 1*10-11 Torr  to 1*104 Torr
• RS-232/RS-485/USB digital communications
• Control up to 10 Televac® vacuum gauges
• Easy-to-read OLED display

Televac®️ MM200 Vacuum Controller

MM200 Vacuum Controller

• Customer configurable design
• RS-232/RS-485 digital communications
• Supports up to 8 relay set points
• Selectable units of measurement